Abstract
Background/Aims: Because of a high monitoring demand and an ensuing need for automation of regional citrate anticoagulation (RCA), a new semi-automated target-oriented algorithm was developed. The aim of this study was the evaluation of its functionality and safety. Methods: Fourteen haemodialysis patients were treated 5 times consecutively with RCA. Samples were drawn pre- and post-infusion once per hour. Electrolytes, blood cell counts, acid-base and coagulation parameters were analyzed. Results: Mean ionized calcium (Ca<sup>2+</sup>) values pre-filter were 0.23 and 0.33 mmol/l in the 0.2 and 0.3 mmol/l target groups, respectively. Extraction ratios for citrate and total calcium through the dialysis filter were constant during the entire treatment (83 and 68%, respectively). Citrate accumulation was avoided. Conclusion: The new algorithm enables safe and accurate RCA. By regulating Ca<sup>2+</sup> pre-filter using the target-oriented algorithm, the degree of anticoagulation may be easily controlled.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.