Abstract

BackgroundSoft tissue injuries are frequently repaired using various suture material. The ideal suture should have the biomechanical properties of low displacement, high maximum load to failure, and high stiffness to avoid deformation. Since tendon healing occurs over a period of months, it is important for the surgeon to select the proper suture with certain biomechanical properties. Therefore, the purpose of this study is to qualitative summarize the published literature on biomechanical properties of different suture materials used in orthopaedic procedures. MethodsFollowing PRISMA guidelines, PubMed and Cochrane databases were queried for original articles containing “biomechanic(s)” and “suture” keywords. Following screening for inclusion and exclusion, final articles were reviewed for relevant data and collected for qualitative analysis. Data collected from each study included the tissue type repaired, suture material, and biomechanical properties, such as elongation, maximum load to failure, stiffness, and method of failure. Results17 articles met final inclusion criteria. Two studies found No.2 Fiberwire™ to have the lowest elongation and 4 studies found No. 2 Ultrabraid™ to have the greatest. 12 studies reported Maximum load to failure was highest in No. 2 Fiberwire™, No. 2 Ultrabraid™, and FiberTape™ while No. 2 Ethibond ™ had the lowest in 5 studies. 3 of the 5 studies that evaluated No. 2 Fiberwire™ found it to have the highest stiffness. No. 2 Ethibond™, No. 2 Orthocord™, and No. 2 PDS™ were reported as the least stiff sutures in 2 studies each. ConclusionFiberwire™, FiberTape™, and Ultrabraid™ demonstrated the highest load to failure while Ethibond™ consistently was the weakest. Fiberwire™ was found to have the lowest elongation while Ultrabraid™ had the highest. Fiberwire™ was also noted to be the stiffest while PDS, Ethibond™, and Orthocord™ were found to be the least stiff. Final treatment decisions on which suture to utilize to optimize repair integrity and healing are complex, and rarely solely dependent upon the biomechanical properties of the materials used. Level of evidenceSystematic Review, Level IV.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call