Abstract

A two-chord, four-beam suprathermal electron diagnostic has been installed on TdeV (B>1.5 T, R=0.86 m, a=0.25 m). Resonant absorption of extraordinary mode electron cyclotron waves is measured to deduce the chordal averaged suprathermal electron distribution function amplitude at the resonant momentum. Simultaneously counterpropagating beams permit good refractive loss cancellation. A nonlinear frequency sweep leads to a concentration of appropriately propagating power in a narrow range of time of flight, thus increasing the signal-to-noise ratio and facilitating the rejection of spurious reflections. Numerous measurements of electron distribution functions have been obtained during lower-hybrid current-drive experiments.