Abstract

Medical gases can be identified simply if they are supplied at different pipeline pressures. Routine maintenance errors, accidental pipeline cross-connections, or gas failures, can be automatically detected by incorporating a pneumatic unit with pressure-sensitive valves into the anaesthetic machine. A fail-safe method of ensuring the delivery of the correct gases to the patient is described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.