Abstract

Past efforts at routine evaluation of gamma camera systems performance, using final product images, primarily resulted in qualitative assessments. This paper describes a system for quantitatively determining certain performance parameters of gamma camera systems using paper-backed print and transparent film test images of the same type used for viewing by the nuclear medicine physician. By use of appropriate transmission and line source phantoms, the gamma camera system's spatial resolution, distortion, and flood field uniformity as a function of position in the image field, are determined using a semiautomated image scanning and data reduction system. This system consists of a high-resolution scanning microdensitometer/reflectometer, interfaced to a computer, along with necessary programs to locate bar and line grid intersection points in the image. Also included are algorithms for calculating the various image parameters desired. This system may provide an analytical method for quantitatively assessing certain performance parameters of gamma camera systems, using the final-product images from the cameras.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.