Abstract

A technique to measure total body calcium using the 40Ca(n, alpha )37Ar reaction has been developed. The technique is based on collecting 37Ar exhaled in the breath following a 10 mrad uniform total body irradiation by 14 MeV neutrons. The 37Ar in the exhaled breath is extracted by selective absorption and its radioactivity is measured inside a low-background proportional detector. The facilities developed include an activation facility providing a +or-2.7% activation uniformity, a closed circuit rebreathing and gas collection system, and a gas purification and counting system. The technique provides a precision of +or-2.4% as determined by repetitive measurements of human volunteers and has an accuracy of determination of total body calcium in grams of +or-5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.