Abstract

A simple, rapid method for determining dissolved inorganic carbon in water is described. A 20-cm3 sample of water is drawn into a 50-cm3 polypropylene syringe and acidified by injection of 1 cm3 of dilute sulphuric acid. Twenty-nine cubic centimeters of helium at atmospheric pressure is injected into the syringe followed by 10 sec of manual agitation to partition CO2 between gas and liquid phase. The gas phase containing 60% of CO2 from the sample is then analyzed by gas chromatography. This method has been used to determine dissolved inorganic and organic carbon in Canadian Shield waters and to determine total carbonates in sediments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.