Abstract
As the scan speed of the atomic force microscope (AFM) operating in intermittent contact mode is increased, the likelihood of artifacts appearing in the image is increased due to the probe tip losing contact with the sample. This paper presents an analysis of the effects of probe loss and a new method, switched gain resonant control, of reducing the problem of probe loss when imaging at high speed. The switched gain resonant controller is implemented to switch the cantilever quality Q factor according to the sample profile during the scan. If the controller detects that the probe has lost contact with the sample the cantilever Q factor is increased leading to a faster response of the feedback controller, expediting the resumption of contact. A significant reduction in image artifacts due to probe loss is observed when this control technique is employed at high scan speeds.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.