Quality assurance (QA) systems for intensity‐modulated radiation therapy (IMRT) have become standard tools in modern clinical medical physics departments. However, because formalized industry standards or recommendations from professional societies have yet to be defined, methods of IMRT QA analysis vary from institution to institution. Understanding where matters stand today is an important step toward improving the effectiveness of IMRT QA and developing standards.We therefore conducted an IMRT QA survey. This particular survey was limited to users of an electronic two‐dimensional diode array device, but we took care to keep the questions as general and useful as possible. The online survey polled institutions (one survey per institution) on a collection of questions about methods of IMRT QA. The topics were general to the IMRT QA analysis methods common to all IMRT systems; none of the questions was vendor‐ or product‐specific.Survey results showed that a significant proportion of responding institutions (32.8%) use the single‐gantry‐angle composite method for IMRT QA analysis instead of field‐by‐field analysis. Most institutions perform absolute dose comparisons rather than relative dose comparisons, with the 3% criterion being used most often for the percentage difference analysis, and the 3 mm criterion for distance‐to‐agreement analysis. The most prevalent standard for acceptance testing is the combined 3% and 3 mm criteria. A significant percentage of responding institutions report not yet having standard benchmarks for acceptance testing—specifically, 26.6%, 35.3%, and 67.6% had not yet established standard acceptance criteria for prostate, head and neck, and breast IMRT respectively.This survey helps in understanding how institutions perform IMRT QA analysis today. This understanding will help to move institutions toward more standardized acceptance testing. But before standards are defined, it would be useful to connect the conventional planar QA analyses to their resulting impact on the overall plan, using clinically relevant metrics (such as estimated deviations in dose–volume histograms).PACS numbers: 87.50.Gi, 87.52.Df, 87.52.Px, 87.53.Dq, 87.53.Tf, 87.53.Kn, 87.56.Fc

Highlights

  • 77 Nelms and Simon: A survey on quality assurance analysis...sometimes used in conjunction with a simple intensity-modulating device such as a solid or electronic wedge

  • The first goal was achieved by framing the topics around the general categories and the tools and techniques that are common to all the major intensity-modulated radiation therapy (IMRT) quality assurance (QA) systems

  • Survey results: general methods Among the 64 follow-up respondents, 64.1% said that they use field-by-field exposure and analysis all of the time and never use the single-gantry-angle composite (SGAC) technique; 32.8% said that they use SGAC technique for more than 75% of all IMRT plans

Read more

Summary

Introduction

Sometimes used in conjunction with a simple intensity-modulating device such as a solid or electronic wedge In this simpler context of three-dimensional (3D) conformal therapy, the medical physicist could, for example, use proper quality control of the aperture manufacturing process, multileaf collimator (MLC) calibration, and careful modeling of dose estimation by the treatment planning system (TPS) to gain the confidence necessary to approve plan accuracy without the need for QA measurement of each and every beam. Despite the development of new delivery and planning tools, the development of efficient and thorough IMRT QA tools lagged slightly at first, leaving medical physicists to do their work with the limited tools available to them. These new clinical and practical needs created a niche for commercial IMRT QA products, and subsequently a set of IMRT QA systems emerged

Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call