Abstract

In order to set up a successful mammography screening program in the Republic of Srpska, a Siemens Mammomat 1000 X-ray machine was selected for analysis as the said mammography system is widely used in clinical practice. The variations in tube parameters (specific air kerma, high-voltage accuracy and reproducibility, linearity between exposure and dose exposure time) were monitored over a five-year period, from 2008 to 2012. In addition, due to observed daily fluctuations for chosen parameters, a series of measurements were performed three times a day within a single-month period (mainly October 2012). The goal of such an experimental set up is to assess short-term and long-term stability of tube parameters in the given mammography unit and to make a comparison between them. The present paper shows how an early detection of significant parameter fluctuations can help eliminate irregularities and optimize the performance of mammography systems.