Abstract

A gravity gradiometer designed for operation on an aircraft or ship has been tested in the laboratory. A noise level of 0.53 E (E ≡ 10(-9) s(-2)) rms over a 0.001 to 1 Hz bandwidth has been measured, and the primary error mechanisms have been analyzed and quantified. The design is a continuation in the development of superconducting accelerometer technology at the University of Maryland over more than three decades. A cryogenic instrument presents not only the benefit of reduced thermal noise, but also, the extraordinary stability of superconducting circuits and material properties at very low temperatures. This stability allows precise matching of scale factors and accurate rejection of dynamic errors. The design of the instrument incorporates a number of additional features that further enhance performance in a dynamically noisy environment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.