Abstract
An exposure system adapted for use on a microscope stage was constructed for studying the effects of high electric field, subnanosecond pulses on biological cells. The system has a bandpass of 3GHz and is capable of delivering high-voltage electric pulses (6.2kV) to the electrodes, which are two tungsten rods (100μm in diameter) in parallel with a gap distance of 170μm. Electric pulses are delivered to the electrodes through a π network, which serves as an attenuator as well as an impedance matching unit to absorb the reflection at the electrodes. By minimizing the inductance of the pulse delivery system, it was possible to generate electric fields of up to 200kV/cm with a pulse duration of 500ps at the surface of the cover slip under the microscope. The electric field at the cover slip was found to be homogenous over an area of 50-70μm. Within this area, neuroblastoma cells placed on the cover slip were studied with respect to membrane potential changes caused by subnanosecond pulses. This allowed us, for the first time, to demonstrate depolarization of the cell membrane potential.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Medical & Biological Engineering & Computing
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.