Abstract

A microprobe system has been installed on the nanoprobe/XAFS beamline (BL8C) at PLS-II, South Korea. Owing to the reproducible switch of the gap of the in-vacuum undulator (IVU), the intense and brilliant hard X-ray beam of an IVU can be used in X-ray fluorescence (XRF) and X-ray absorption fine-structure (XAFS) experiments. For high-spatial-resolution microprobe experiments a Kirkpatrick-Baez mirror system has been used to focus the millimeter-sized X-ray beam to a micrometer-sized beam. The performance of this system was examined by a combination of micro-XRF imaging and micro-XAFS of a beetle wing. These results indicate that the microprobe system of the BL8C can be used to obtain the distributions of trace elements and chemical and structural information of complex materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.