Abstract

Organic scintillator detectors are widely used for neutron spectroscopy in fusion devices due to their good energy resolution and capability of neutron/gamma discrimination. Nowadays, scintillator pulses are commonly recorded by means of digital acquisition systems. These have several advantages, and in particular the possibility of off-line data reprocessing: however, the signal digitization can be a cause of degradation of the pulse height (PH) resolution (and therefore of the energy resolution obtained after pulse height spectra unfolding). In this work, the problem of how pulse digitization may influence the pulse height resolution is investigated. First, through downsampling of digitized scintillator pulses, we determine the minimum sampling rate required to avoid any degradation of the pulse height resolution. Secondly, we find that the application of interpolation methods to the digitized pulses does not affect the pulse height resolution, whereas it increases the figure of merit for neutron/gamma discrimination. These results are relevant to define the specifications for the digital acquisition systems of neutron detectors in present and future fusion devices such as JET and ITER.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.