Abstract
This study was designed to compare the results of hearing tests performed using the uHear application with those of standard audiometry in Thai people in Bangkok. From December 2018 to November 2019, a prospective observational study was conducted involving Thai participants aged between 18 and 80 years. All participants were tested using standard audiometry and the uHear application in a soundproof booth and in a typical hearing environment. This study included 52 participants (12 males and 40 females). The Bland-Altman plot with the Minimal Clinical Meaningful Difference of 10dB between standard audiometry and the uHear in a soundproof booth found agreement at 2000 Hz. The uHear in a soundproof booth showed high sensitivity at all frequencies (82.5%-98.9%) and high specificity at 500 and 1000 Hz (85.7%-100%). uHear in a typical hearing environment showed high sensitivity at 4000 and 6000 Hz (97.6%) and high specificity at 500 and 1000 Hz (100%). When considering the pure-tone average, uHear in a soundproof booth showed high sensitivity (94.7%) and specificity (90.7%), whereas, in a typical hearing environment, uHear showed poor sensitivity (34%) and high specificity (100%). uHear was accurate for hearing loss screening at 2000 Hz in a soundproof booth. However, uHear in a typical hearing environment lacked accuracy. The uHear application in a soundproof booth can be used to screen hearing loss in some situations where standard audiometry is impossible. II.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.