Abstract
Experiments were made with a vortex meter situated downstream of two elbows out-of-plane, at Reynolds numbers between 7.8×10 4 and 2.1×10 5. Analysis on the signals obtained from a piezo-electric sensor embedded in the vortex shedder shows that the signal quality, concerned with the vortex shedding frequency component, deduced in the vicinity of 15 pipe diameters downstream is superior to that at other locations, and is remarkably better than that found in the region of fully-developed turbulent pipe flow. The vortex shedding frequencies measured in this region, in terms of the Strouhal value, fall within 0.129±0.001, that 0.129 is the reference value reduced in the fully-developed turbulent pipe flow region. Also noted, the uncertainty interval associated with the vortex shedding frequency measured in this region is smaller than that reduced in the region of fully-developed turbulent pipe flow.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.