Abstract

An ion selective electrode (ISE) for determining Cr(VI) using supported liquid membranes (SLMs) containing trioctylphosphine oxide (TOPO) was investigated in this study. TOPO, as a carrier, had a high selectivity for Cr(VI) against interfering ions such as sulfate, nitrate, nitrite, and chloride. The composition of the SLM was optimized as 0.193 g TOPO/1 mL NPOE (o-nitrophenyl octyl ether)/0.5 g poly (vinyl chloride) for detection of Cr(VI). The Cr(VI) concentration was measured in the range of 1 x 10(-3) to 1 x 10(-6) M with the SLM prepared in the study. It seemed that Cr(VI) was transported in the SLM as a triply charged ion indicated by the slope of the emf response. Selectivity coefficients and detection limits of Cr(VI) in the presence of interfering ions were determined experimentally using the fixed interference method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.