Abstract

The quality of the present day fluoroscopic images is sufficiently high for use as exposure images depending on the environment where the fluoroscopic images are recorded. In some facilities which use fluoroscopic images as exposure images they are recorded with a radiological x‐ray diagnostic device equipped with a fluoroscopic storage function. There are, however, cases where fluoroscopic images cannot be used as exposure images because the quality of the fluoroscopic image cannot be assured in the environment where the fluoroscopic images are recorded. This poses problems when stored fluoroscopic images are used in place of exposure images without any clearly established standard. In the present study, we establish that stored fluoroscopic images can be used as exposure images by using gray values obtained from profile curves. This study finds that replacement of stored fluoroscopic images with exposure images requires 20.1 or higher gray scale value differences between the background and signal, using a 20 cm thick acrylic phantom (here an adult abdomen as representing the human body) as the specific geometry. This suggests the conclusion that the gray value can be considered a useful index when using stored fluoroscopic images as exposure images.