Abstract

An experimental study was conducted to document system head losses for dental vacuum systems. Experiments were performed to estimate pressure losses and pressure loss coefficients for pertinent dental treatment room hoses and fittings in high-volume evacuation systems. Pressure losses for some polyvinyl chloride (PVC) elbows used in dental plumbing systems were also measured and presented. Friction head losses in straight PVC piping are calculated and documented in graphical form. Calculated values for friction losses are found to correlate exceptionally well with measured experimental data.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.