Abstract

Objectives Pre-analytical error decisively influences the total error and consequently the diagnostic accuracy. The following were the objectives for the study: 1. To detect the percent of pre-analytical errors in clinical biochemistry laboratory 2. To categorize these pre-analytical errors 3. To formulate corrective measures to be taken to avoid such errors Design and methods Study period was for 3 months with documenting the frequency and type of pre-analytical errors occurring in the venous samples. Result Average pre-analytical errors were 44.7% per day. Improper request, incorrect timing of sample, wrong tube collection and in-vitro hemolysis of samples amounted to the major proportion of errors. Conclusion Pre-analytical errors occurring in each laboratory have to be checked. Such errors are not inevitable and can be avoided with a diligent application of quality control, continuing education and effective collection systems to ensure total quality patient care.