Medical accelerators with photon energies over 10MeV generate an undesired fast neutron contamination in the therapeutic beam. In this work, the Monte Carlo code MCNP was used to simulate the transport of these photoneutrons across the head of various medical accelerators of high energy. The average and most probable neutron energies were obtained from these spectra, before and after crossing the accelerator shielding. The degradation of these spectra, when they cross concrete barriers with thickness which vary between 25 and 100cm, was also studied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.