Abstract
Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have been obtained experimentally. Measurement results are compared with that of the Monte Carlo calculations reported, and no difference has been found between them.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.