Abstract

Chemically treated, biologically derived soft collagenous tissues are used extensively in medical devices. To enable prosthesis design through computational methods, physically realistic constitutive models are required. In the present study, a structural approach was utilized that incorporated experimentally measured angular distribution of collagen fibers. Using biaxial mechanical data from our previous study (Annals of Biomedical Engineering, vol. 26(5), pp. 892–902, 1998), the effective fiber and matrix stress–strain responses were predicted. The agreement with the experimental data supported the assumption that the mechanical effects of chemical treatment are equivalent to the addition of an isotropic elastic matrix. An important utility of this model is its ability to separate the effects of chemical treatment on the fibers and matrix. Applications of this approach include utilization in the design of novel chemical treatments that produce specific mechanical responses, the study of fatigue damage, and finite element implementation for tissue engineering scaffold design.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.