Abstract

In this paper, we present a stretchable electrode array for studying cell behavior subjected to mechanical strain. The electrode array consists of four gold nail-head pins (250?m tip diameter and 1.75mm spacing) inserted into a polydimethylsiloxane (PDMS) platform (25.4×25.4mm2). Fusible indium alloy (liquid at room temperature) filled microchannels are used to connect the electrodes to the outside, thus providing the required stretchability. The electrode platform is biocompatible and can withstand strains of up to 40%. We tested these electrodes by repeatedly (100 times) subjecting them to 35% strain and did not notice any failure. We also successfully cultured mice cardiomyocytes onto the platform and performed electrical pacing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.