Abstract

Frequency-dependent phase velocity (dispersion) has previously been measured in trabecular bone by several groups. In contrast to most biologic tissues, phase velocity in trabecular bone tends to decrease with frequency. A stratified model, consisting of alternating layers of bone and marrow (in vivo) or water (in vitro), has been employed in an attempt to explain this phenomenon. Frequency-dependent phase velocity was measured from 300 to 700 kHz in 1) phantoms consisting of regularly spaced thin parallel layers of polystyrene sheets in water and 2) 30 calcaneus samples in vitro. For the polystyrene phantoms, the agreement between theory and experiment was good. For the calcaneus samples, the model has some limited usefulness (uncertainty of about 5%) in predicting average phase velocity. More importantly, the model seems to perform consistently well for predicting the frequency dependence of phase velocity in calcaneus.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.