Abstract

Knowledge of the central-axis depth-dose curve and the dose profiles of a radiotherapy treatment unit are essential for setting up an external system for planning radiotherapy treatment. Estimates for these are usually found with the aid of a measurement probe with which the dose is measured at equally spaced measurement points. In this paper a strategy for selecting the measurement points for determination of the depth-dose curve and dose profile is proposed. The performance of the method is evaluated with real 6 MV photon depth-dose data and real 18 MV dose-profile data. The studies show that the proposed method allows the estimation of the depth-dose curve with adequate accuracy typically from 10 - 14 measurements at this energy instead of approximately 120 measurements, which is a common number of measurements. Thus the measurements protocol decreases about tenfold the number of measurements required. The studies also show that a fivefold decrease can be achieved in the dose-profile measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.