Abstract
Combining stimulated emission depletion and fluorescence correlation spectroscopy (STED-FCS) provides a powerful and sensitive tool for studying the molecular dynamics in live cells with high spatio-temporal resolution. STED-FCS gives access to molecular diffusion characteristic at the nanoscale occurring within short period of times. However due to the incomplete suppression of fluorescence in the STED process, the STED-FCS point spread function (PSF) deviates from a Gaussian shape and challenges the analysis of the auto-correlation curves obtained by FCS. Here, we model the effect of the incomplete fluorescence suppression in STED-FCS experiments and propose a new fitting model improving the accuracy of the diffusion times and average molecule numbers measurements. The implementation of a STED module with pulsed laser source on a commercial confocal/FCS microscope allowed us to apply the STED-background corrected model to fit the STED-FCS measurements. The experimental results are in good accordance with the theoretical analysis both for the number of molecules and the diffusion time which decrease accordingly with the STED power.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.