Abstract

The dried droplet method is proposed as a way of characterizing matrix effects in LIBS with a standardized approach. This method was introduced first in the field of LA-ICPMS for quantitative analysis of solids. It consists in depositing a droplet of an iron-containing solution on the sample surface and ablating the dry residue. Then, iron lines are used for spectroscopic diagnoses of the plasma. Along with white-light profilometry analysis for ablation craters measurements, this aims to accurately determine differences of ablated mass, electron excitation temperature and number density, including for pure metals. In this paper, we check that the presence of the dry residue does not influence those three factors. Then, the dried droplet method is applied to 14 pure metals. Results show that the number of ablated atoms varies by a factor of 25, while for the electron excitation temperature the maximum gap between the 14 metals is approximately 2000 K, i.e. a relative variation of ~30%. As for the electron number density, it could be estimated for only 6 metals and it varies within the measurement uncertainty between 6 and 8 1016 cm−3.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.