Abstract

We describe a method to perform noncontact atomic force microscopy (NC-AFM) in liquid and in air using a quartz crystal tuning fork. The latter, thanks to its increased stability originating from its higher stiffness, compared to cantilevers, enables very low oscillation amplitudes and thus, should yield a very high sensitivity to the force interactions. However, applying conventional NC-AFM technique fails, due to the phase lock loop demodulator becoming unstable as the input signal to noise ratio decreases when lowering the oscillation amplitude. Therefore we have developed so-called zero phase mode atomic force microscopy based on phase modulation. Comparisons with conventional frequency modulation mode using the same mechanical setup demonstrate gains in sensitivity and in scanning rates. High resolution topographic images, as well as simultaneous dissipation images can be recorded using this very convenient and effective technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.