Abstract

We proposed a spherical coil array for determining the positions, orientations, and sensitivities of the magnetometers for whole-head magnetoencephalograph (MEG) systems. The coil array comprises 16 concentric and symmetrically oriented circular coils with a diameter of 150 mm, which is larger than those of conventional calibration coil arrays, such as 5 mm or 60 mm. The accuracy of the calibration was expected to improve because the relative mechanical errors in machining and assembly could be reduced when the coil diameter was increased. Moreover, we also proposed a method to estimate the uncertainties in calibration, and it was demonstrated that the sensor parameters of each sensor were successfully obtained with uncertainties using a spherical calibration coil array. Additionally, the accuracy of the calibration was evaluated using a dry-type MEG phantom, and as expected, it was found that the accuracy was improved from that of a conventional calibration coil array composed of an assembly of multiple 3-axis bobbins.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.