Abstract

The purpose of this investigation was to determine the strength and mode of failure of a plate specifically designed for wrist arthrodesis. An eight-hole plate was constructed with a solid central portion and a distal taper. Four 2.7 mm screws were used in the distal (metacarpal) portion of the plate. Four 3.5 mm screws were used in the proximal (radius) portion. The strength of the wrist arthrodesis plate was compared with that of a conventional 10-hole 3.5 dynamic compression plate by gap testing and a paired cadaver wrist arthrodesis construct. The strength of the wrist arthrodesis plate was similar to the strength of the conventional dynamic compression plate when applied to cadaver wrist arthrodesis constructs. Failure of the wrist arthrodesis constructs occurred through bone without deformation of either plate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.