Abstract

90Sr+90Y applicators are commonly utilized in brachytherapy, including ophthalmic procedures. The recommended instruments for the calibration of these applicators are extrapolation chambers, which are ionization chambers that allow the variation of their sensitive volume. Using the extrapolation method, the absorbed dose rate at the applicator surface can be determined. The aim of the present work was to develop a mini-extrapolation chamber for the calibration of 90Sr+90Y beta ray applicators. The developed mini-chamber has a 3.0 cm outer diameter and is 11.3 cm in length. An aluminized polyester foil is used as the entrance window while the collecting electrode is made of graphited polymethylmethacrylate. This mini-chamber was tested in 90Sr+90Y radiation beams from a beta particle check source and with a plane ophthalmic applicator, showing adequate results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.