Abstract

We have improved the sonic spray interface to enable the analysis of multiply charged ions of protein from a solution at a flow rate of 1 mL/min using a conventional liquid chromatograph/mass spectrometer. In this interface, we added a multihole plate in front of the sampling orifice of a mass spectrometer. This plate does not have a hole coaxial to the sampling orifice but has small holes around the central region of the plate. The plate reduces the density of the solvent molecules in the sprayed gas introduced into the vacuum region through the sampling orifice from the atmosphere and prevents the ions from being solvated and becoming charged droplets due to the cooling that follows adiabatic expansion of the sprayed gas. With this improvement, multiply charged ions whose charge distribution ranged from 11+ to 16+ were analyzed from a 1 μM cytochrome c solution at a high flow rate of 1 mL/min without using a splitter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.