Abstract

BackgroundMeasurement of carbohydrate antigen (CA) 19-9 is not applicable in patients with Lewis (Le) blood type, Le(a−b−). It is important to distinguish cases with Le(a−b−) before CA19-9 measurement. Therefore, we prepared a cut-off solution that gives a clear index to distinguish Le(a−b−) sera. MethodThe frequencies of Le blood types and the distribution of the CA19-9 values in each Le blood type were examined in 188 healthy subjects. The CA19-9 values for all Le(a−b−) sera and for a portion of Le(a−b+) sera exist below the limit of quantitation as measured by the SphereLight 180 kit. The cut-off solution, which gives a clear cut-off index (COI), was prepared to differentiate Le(a−b−) from Le(a−b+), and was evaluated using the SphereLight 180, Architect i2000, UniCel DxI 800, Elecsys 2010, and Lumipuls ƒ kits. ResultsThe COI was calculated as the mean +3 SD of the CA19-9 values of a cut-off solution that is adjusted to the limit of detection. Both the sensitivities and specificities of the COIs were 100% using the SphereLight 180 kit and 100% and 91.7%, respectively, using the Architect i2000 kit, but these values were not satisfactory using the other CA19-9 assay kits. ConclusionThe COIs, determined by the cut-off solution, correctly identified all Le(a−b−) sera as Le(a−b−) and differentiated Le(a−b−) sera from other types of sera in CA19-9 assays using only the SphereLight 180 and Architect i2000 kits.