Abstract

We have built an electrostatic electron spectrometer combining both high resolution and large luminosity. The instrument consists essentially of two superconducting solenoids separated by a system of ring electrodes of 4 m in length. Source and detector are placed in the high-field regions of the superconducting solenoids, whereas the repellent analyzing electrostatic potential of the ring electrodes peaks at the minimum of the magnetic field in between these solenoids. The magnetic guiding field provides (i) the acceptance of the full foreward solid angle of 2π, (ii) the transformation of the transverse cyclotron motion into longitudinal motion parallel to the magnetic field. The energy resolution of the electrostatic filter is determined by the ratio of the magnetic fields at the source and in the analyzing plane. It is typically 5 × 10 3 in our case. The spectrometer will serve first of all to investigate the limits of the rest mass of the electron antineutrino from 3H 2ß-decay. It has been tested by measuring conversion lines from a 83mKr source which yielded an energy of Eγ = 32151.5(11) eV for the corresponding nuclear transition.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.