Abstract
The design of a tissue slice perfusion system is described, and examples are given showing the stability of this system for intracellular and extracellular recordings during changes in perfusion media. The stability of this system is attributed to several features. Mini-drips serve to cushion transient changes in flow rate when switching from one medium to another. Solenoid valves are used to quickly switch perfusion media with minimal mechanical movement. A finely-controlled adjustable flow valve provides a uniform flow rate for all media. Constant tissue temperature is maintained by media perfusion through a thermoelectric Peltier assembly. In addition, a filter paper wick insures that the perfusate is constantly removed without movement in the tissue slices. With this design, the slices are supported on a net at the interface between the perfusion medium and a humidified, oxygenated atmosphere. This arrangement appears to be conducive to tissue viability and facilitates the placement of microelectrodes in the slices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.