Abstract

The limitations of traditional ureteral stents in patients with deficiencies in ureteral drainage have resulted in frequent stent exchanges. The implementation of metallic stents was introduced to improve the patency rates of patients with chronic upper urinary tract obstruction, obviating the need for frequent stent exchanges. We report our clinical experiences with the use of metallic ureteral stents in the management of poor ureteral drainage. Fifty patients underwent metallic ureteral stent placement from 2009 to 2012. Stent failure was defined as an unplanned stent exchange, need for nephrostomy tube placement, increasing hydronephrosis with stent in place, or an elevation in serum creatinine. Stent life was analyzed using the Kaplan-Meier methodology, as this was a time dependent continuous variable. A cost analysis was similarly conducted. A total of 97 metallic stents were placed among our cohort of patients: 63 in cases of malignant obstruction, 33 in the setting of cutaneous ureterostomies, and 1 in an ileal conduit urinary diversion. Overall, stent failure occurred in 8.2% of the stents placed. Median stent life was 288.4 days (95% CI: 277.4-321.2 days). The estimated annual cost for traditional polymer stents (exchanged every 90 days) was $9,648-$13,128, while the estimated cost for metallic stents was $4,211-$5,313. Our results indicate that metallic ureteral stent placement is a technically feasible procedure with minimal complications and is well tolerated among patients. Metallic stents can be left in situ for longer durations and provide a significant financial benefit when compared to traditional polymer stents.

Highlights

  • The limitations of traditional ureteral stents in patients with deficiencies in ureteral drainage have resulted in frequent stent exchanges

  • A total of 97 metallic stents were placed in 50 patients (27 men, 23 women) during the 45-month accrual period

  • The mean patient age at diagnosis was 63.0 years (22-88 years). 37 patients (74%) had stents placed due to malignant ureteral obstruction, 12 patients (24%) had stents placed in the setting of cutaneous ureterostomies, and 1 patient (2%) had stent placement in the context of an ileal conduit. 14 patients (28%) had stents placed for genitourinary malignancies, 7 patients (14%) had stents placed for gastrointestinal malignancies, and 16 patients (32%) had stents placed for other malignancies, including lymphoma and sarcoma

Read more

Summary

Introduction

The limitations of traditional ureteral stents in patients with deficiencies in ureteral drainage have resulted in frequent stent exchanges. The implementation of metallic stents was introduced to improve the patency rates of patients with chronic upper urinary tract obstruction, obviating the need for frequent stent exchanges. Stent failure was defined as an unplanned stent exchange, need for nephrostomy tube placement, increasing hydronephrosis with stent in place, or an elevation in serum creatinine. Results: A total of 97 metallic stents were placed among our cohort of patients: 63 in cases of malignant obstruction, 33 in the setting of cutaneous ureterostomies, and 1 in an ileal conduit urinary diversion. Conclusion: Our results indicate that metallic ureteral stent placement is a technically feasible procedure with minimal complications and is well tolerated among patients. Standard of care recommends regular stent replacement at 3-month intervals to prevent failure [3,4]

Methods
Results
Conclusion