Abstract

The assessment of low-contrast-details is a part of the quality control (QC) program in digital radiology. It generally consists of evaluating the threshold contrast (Cth) detectability details for different-sized inserts, appropriately located in dedicated QC test tools. This work aims to propose a simplified method, based on a statistical model approach for threshold contrast estimation, suitable for different modalities in digital radiology. A home-madelow-contrast phantom, made of a central aluminium insert with a step-wedge, was assembled and tested. The reliability and robustness of the method were investigated for Mammography, Digital Radiography, Fluoroscopy and Angiography. Imageswere analysed using our dedicated software developed on Matlab®. TheCth is expressed in the same unit (mmAl) for all studied modalities. This method allows the collection of Cthinformation from different modalities and equipment by different vendors, and it could be used to define typical values. Results are summarized in detail. For 0.5 diameter detail, Cthresults are in the range of: 0.018-0.023mmAl for 2D mammography and 0.26-0.34mmAl DR images. For angiographic images, for 2.5mm diameter detail, the Cths median values are 0.55, 0.4, 0.06, 0.12mmAl for low dose fluoroscopy, coronary fluorography, cerebral and abdominal DSA, respectively. The statistical method proposed in this study gives a simple approach for Low-Contrast-Details assessment, and the typical values proposed can be implemented in a QA program for digital radiology modalities.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call