Abstract

We present the design and performance of a broad-band single-chip integrated transceiver specifically conceived for nuclear magnetic resonance magnetometry. The single-chip transceiver is realized using a standard silicon complementary metal-oxide-semiconductor integrated circuit technology. A radio-frequency (RF) transmit amplifier, a transmit/receive switch, a low noise RF receive amplifier, a quadrature (IQ)-mixer, and two intermediate frequency amplifiers are integrated on a single silicon chip of 1.8 mm2. The advantages and problematic aspects with respect to conventional discrete electronic approaches are discussed. We show the results of magnetic field measurements performed at 1.4 and 7.05 T, using solid and liquid samples having volumes from 40 μl down to 100 pl. Particular attention is devoted to the comparison of the experimentally measured magnetic field standard deviation with respect to the Cramer-Rao lower bound value. With a sample of distilled water (T1 ≅ T2 ≅ 3 s, T2 *≅ 20 ms) having a volume of 40 μl, a standard deviation of 2.5 nT at 7.05 T (i.e., 0.5 ppb) in 1 s of averaging time is achieved, with a projected Cramer-Rao lower bond of 8 pT (i.e., 1.1 ppt).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.