Abstract

By modifying the wavelength of the open path tunable diode laser spectrometer (TDLS) in the near infrared region, theoretical research was implemented to improve the detection limit of carbon monoxide gas. To adjust the correct wavelength in the NIR area, MatLab code was created. Following that, frequency domain measurements were performed in order to extract the second harmonic as an indicator of gas presence. According to the results, the correct wave length in the NIR area is (1584.877 nm), and the lowest limit of CO gas concentration is (0.012 ppb).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.