Abstract

We proposed and evaluated a water-equivalent diameter calculation without using a region of interest (ROI), (Dw,t) and compared it with the results of using a ROI fitted to the patient border (Dw,f). Evaluations were carried out on thoracic and head CT images. We found that the difference between Dw,t and Dw,f was within 5% for all images in the head region, and most images were within 5% (27 of the 30 patients, 90%) in the thoracic region. We also proposed a method to automatically detect and eliminate the patient table (or head support) from images and evaluated the water-equivalent diameter values after the table had been removed (Dw,nt). This method was able to recognize and remove the patient table from all images used. By removing the table, the water-equivalent diameter (Dw,nt) became more accurate and the difference from Dw,f was within 5% for all images (head and thoracic images).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.