Abstract

We have developed and report on a simplified cardiopulmonary bypass technique for experiments on implantable ventricular assist devices in calves. We used an electromechanical implantable ventricular assist device with a double cylindrical cam in three calves. Cannulas for the ventricular assist system were designed to be inserted between the left atrium and the descending aorta. We used the outflow cannula of the ventricular assist device, anastomosed to the descending aorta, as a temporary arterial return route for the cardiopulmonary bypass. A cannula for venous drainage was iserted into the right ventricle through the pulmonary artery. There were no problems related to the procedure and the cardiopulmonary bypass was succesful. In conclusion, this simplified cardiopulmonary bypass technique without neck incision in calves, as used in developmental work involving implantable ventricular assist devices, can be reliably performed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.