A simple algebraic manipulation of known formulas leads to a direct expression for the specific airway resistance (SRaw) which precludes separate measurements of the airway resistance (Raw) and the thoracic gas volume (TGV). The equation is: SRaw = tg beta (PBar - PH2O) e2, in which tg beta stands for the relation between the plethysmographic box volume and pneumotachograph flow fluctuations; PBar - PH2O is the barometric por graphical units. Thanks to this new formula SRaw can now be easily measured with great precision and quickly calculated, even in the case of those children (Table 1) with whom this was not previously possible. No further cooperation is needed than breathing at a normal rate through mouthpiece into a whole-body plethysmograph. The intraindividual coefficient of variation (CV) for SRaw is about 8% and never exceeds 15% (Table 1).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.