Abstract
AbstractA method was developed for the determination of organic halogens in water samples with the aim of minimizing matrix effects and simplifying the sample preparation technique commonly used. The method is based on the adsorption of organic halogens in modified hydrophilic divinylbenzene polymer columns and their elution with methanol. The adsorbent used enables rapid adsorption and desorption due to comparatively high operating flow rates and minimized solvent amounts. Furthermore, no additional reconcentration steps are needed. The methanol extract obtained is combusted with a standard EOX (extractable organic halogen) analyzer and the concentration of organically bound halogens is determined by microcoulometric analysis. The matrix effects are considerably reduced compared to the standard procedure (EN 1485) commonly used. A detection limit of 13 μg/L was established.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.