Abstract
Folic acid (FA) is an ingredient that must be added to infant milk powder to avoid potential defects. Rapid, sensitive and reliable detection methods are needed to determined FA addition levels. Thus, this study established a microsphere immunochromatographic test strip for time-resolved luminescence detection (TRLM-ICTS) based on carboxyl-functionalized time-resolved luminescent microspheres (Eu-TRLMs) prepared by a one-step method as fluorescent markers for the immediate quantitative detection of FA in milk powder. Eu-TRLMs prepared by the one-step method showed good dispersion, high stability and strong fluorescence intensity, which is improving the sensitivity of TRLM-ICTS. In the performance evaluation of TRLM-ICTS, the detection limit was 0.487 ng mL−1, the recovery rate was 97.3–105 %, and the actual sample detection results were in line with those of UPLC-MS/MS. TRLM-ICTS has the advantages of rapid, high sensitivity and strong specificity and could as a practical quantitative detection method for the detection of FA in milk powder.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.