Abstract
The aim of this study was to describe a simple and easy-to-use calibration method that is able to estimate the pose (tip position and orientation) of a rigid endoscopic instrument with respect to an electromagnetic tracking device attached to the handle. A two-step calibration protocol was developed. First, the orientation of the instrument shaft is derived by performing a 360° rotation of the instrument around its shaft using a firmly positioned surgical trocar. Second, the 3D position of the instrument tip is obtained by allowing the tip to come in contact with a planar surface. The results indicate submillimeter accuracy in the estimation of the tooltip position, and subdegree accuracy in the estimation of the shaft orientation, both with respect to a known reference frame. The assets of the proposed method are also highlighted by illustrating an indicative application in the field of augmented reality simulation. The proposed method is simple, inexpensive, does not require employment of special calibration frames, and has potential applications not only in training systems but also in the operating room.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.