Abstract

Cerebral creatine deficiency syndromes (CCDS) are caused by genetic defects in L-arginine:glycine amidinotransferase, guanidinoacetate methyltransferase or creatine transporter 1. CCDS are characterized by abnormal concentrations of urinary creatine (CR), guanidinoacetic acid (GA), or creatinine (CN). In this study, we describe a simple HPLC method to determine the concentrations of CR, GA, and CN using a weak-acid ion chromatography column with a UV detector without any derivatization. CR, GA, and CN were separated clearly with the retention times (mean ± SD, n = 3) of 5.54 ± 0.0035 min for CR, 6.41 ± 0.0079 min for GA, and 13.53 ± 0.046 min for CN. This new method should provide a simple screening test for the diagnosis of CCDS.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call