Abstract

High-quality gradient coils are pivotal to advances in magnetic resonance imaging (MRI). We have studied the influence of coil dimensions and target requirements in multilayer, asymmetric, transverse gradient coils. We developed a simple linear function that defines the optimal coil length to produce a maximum figure of merit given an imaging region size and location, coil radius, and gradient nonuniformity. Our method, based on the linear function, yields high-quality solutions. The method introduces two torque/force minimization strategies in order to obtain asymmetric transverse gradient coils that balance minimum torque with a maximum figure of merit. High-performance head, asymmetric gradient coils with simple current patterns and minimum torque can be tailored to a specific magnet design, as we illustrate

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.