Abstract
A simple, reliable, and innocuous method is described for evaluating the penetrability of cerebrospinal shunts with the use of 99mTc O4. After introducing 25 muCi of 99mTc O4 in a volume of about 0.2 ml into Rickham's reservoir, series scintigrams in the lateral or AP direction are made with the aid of a Nuclear Chicago Pho Gamma HP Scintillation Camera. The results are in satisfactory agreement with the surgical findings. No complications have so far been observed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.