Abstract
The analysis by electrospray-ionization tandem mass spectrometry of amino acids with butyl esterification and isotopically labeled internal standard is routine in newborn screening laboratories worldwide. In the present study, we established a direct analysis method of higher accuracy that uses a non-deuterated internal standard. The automatic sampler and the pump of an LC apparatus were used to inject sample and mobile phase to MS, but no LC column was needed. The dry blood spot (DBS) material was prepared at levels of low, medium and high concentration; the running time was 1min. In parallel to the new procedure, we applied the established method to analyze nine amino acids on DBS of healthy newborns and phenylketonuria newborns. The newly proposed method of product ion confirmation scan along with multiple reaction monitoring resulted in a very accurate identification of each amino acid. Our innovative protocol had high sensitivity and specificity in the analysis of cases of suspected metabolic diseases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.